Abstract
The most common mid-infrared (MIR) attenuated total reflection (ATR) accessory has a nominal angle of incidence of 45° and does not have a polarizer. A spectrum recorded with such an accessory does not hold enough information for the sophisticated ATR correction of MIR spectra with strong peaks, which are often strongly affected by refractive index changes due to anomalous dispersion. Here we show that a 45° ATR spectrum recorded without a polarizer and the polarization angle for the same ATR Fourier transform infrared spectroscopy system provide enough information to determine the ATR s-polarized spectrum. Further analysis with an improved non-iterative Kramers–Kronig analysis immediately yields the complex refractive index function. The analysis is about two orders of magnitude faster than iterative formalism and runs within seconds on a typical office PC. The effectiveness of our advanced ATR correction formalism is showcased through its application to water, employing diamond, ZnSe, and Ge ATR crystals, along with two distinct ATR accessories. Additionally, the formalism is applied to octadecane spectra. Potential sources of errors such as incidence angle spread, dispersion of the polarization angle, and the influence of reflection at the air/ATR crystal interface are investigated by simulations.
This is a visual representation of the abstract.
Introduction
Attenuated total reflection (ATR) accessories are used with many Fourier transform infrared spectroscopy (FT-IR) systems because they require less effort to record a reproducible spectrum than recording a transmission spectrum with KBr pellet preparation. However, ATR introduces changes in peak positions, intensities, and shapes. So-called ATR correction is the two-step process of (1) converting an ATR spectrum to the complex refractive index function,
The differences between ATR and transmission spectra (or reflectance absorbance and transmittance absorbance spectra) go beyond the need for intensity correction, particularly for lower wavenumbers.6–11 The effective thickness, which is the thickness that produces absorbance values equivalent to those obtained from transmission experiments, increases inversely proportional to the wavenumber.
12
Electromagnetic theory correspondingly demonstrates that ATR absorbance depends on the index of refraction and the penetration depth, which is the thickness at which the evanescent wave is reduced to 1/e of its original intensity. The penetration depth is itself influenced by the index of refraction. Consequently, the absorbance band maxima are shifted toward the maxima of the index of refraction, and the band shapes can undergo significant changes.7–11 One way to mitigate this issue is to use high-index ATR crystal materials such as Si and Ge and to employ high angles of incidence.2,13 However, this approach reduces the penetration depths and overall intensities, thereby decreasing the sensitivity of the method. In addition, although effects such as peak shifts and band shape changes are reduced, they remain present to some extent.
2
On the other hand, this approach helps prevent the use of the ATR method outside of its specifications, ensuring that ATR spectra remain internal reflectance spectra. What do these specifications entail? The primary requirement is that the angle of incidence (AOI) must exceed the critical angle, α
c
:2
If, for example, the crystal used in ATR spectroscopy is diamond or zinc selenide (ZnSe), both of which have a refractive index of approximately 2.4, and the AOI α i amounts to 45°, then the refractive index of the sample being analyzed must be <1.7. This means that certain materials, such as charcoal, cannot be analyzed using this combination of crystal and AOI because the resulting spectra are influenced mainly by the refractive index rather than the absorption index. 14 The reason is that according to Fresnel's equations below the critical angle, a fraction of the incident light is transmitted into the sample and the reflectance is therefore generally <1. Accordingly, –log R would be positive, although nothing is absorbed. Additionally, if a fraction of the light that is transmitted into the sample is absorbed, because the absorption index is different from zero, it usually scarcely alters –log R, at least for not-too-strong oscillators (this builds the foundation of infrared [IR] refraction spectroscopy, see Mayerhöfer et al. 14 ). Even when the refractive index of a sample is below the critical angle in the transparency region between mid-IR (MIR) and visible (Vis) light, there can still be problems caused by dispersion. Absorption causes a change in the refractive index of the sample. Therefore, overlapping weak bands or a medium-strong band such as a C=O vibration can result in a spectral range where the critical angle is larger than the incidence angle chosen for an ATR accessory. In this case, some of the radiation in that range is transmitted into the sample without being absorbed. Unfortunately, it is only possible to measure and quantify the deviation from total reflection, but not whether this deviation is solely due to absorption. This means that absorbance values in ranges with a higher critical angle than the actual AOI can seem to be much higher than they actually are.2,11 This is not an issue when spectra are evaluated based on electromagnetic theory, as reflection losses due to transmission are automatically taken properly into account. 2 The only requirement is that the polarization state of the incident light must be known for correction. This is the real challenge at present since most accessories do not have a polarizer.
Theoretical Considerations
Recently, Azam et al.
15
developed a technique for determining the polarization angle by comparing an experimental spectrum of water to a calculated one. By utilizing known optical constants, Fresnel's equations, and Malus’ law, the polarization angle can be computed. According to Malus’ law, the reflectance in dependence of the polarization angle φ is given by:
According to Fresnel's equations, the reflectances Rs and Rp can be calculated according to
2
It can be shown based on Eq. 3 that for an AOI α
i
= 45°, a special relationship between Rs and Rp is established according to which
Note that we have added, compared to the formula provided by Bertie and Lan, a semi-empiric correction factor for the last term, 8002 cm–2/
Depending on the spectral resolution, the calculation requires about 2–10 s on a typical office PC, which means that it is much faster, up to two orders of magnitude, than an iterative correction procedure used, e.g., in Mayerhöfer et al. 20 In contrast to the latter procedure, it usually also works when the condition of total reflection is violated around bands, i.e., when the refractive index of the sample becomes so large that the critical angle is larger than the AOI. For ZnSe and Diamond ATR crystals and an AOI of 45° this happens for refractive indices larger than about 1.7, which is quite often reached, e.g., for C=O or amide I bands. In this case, the so-called “advanced ATR correction” formalisms fail, which are based on a much less sophisticated correction scheme than the one introduced in this work, but do not require knowledge of the polarization angle.
Experimental
Materials and Methods
Two ATR accessories, designated ATR-1 and ATR-2, were used in this work, without temperature controllers. All ATR crystals were single-bounce with 45° faces for the incident and reflected MIR light. ATR-1: A fully automated FDM ExpertATR (Fiveash Data Management, Inc., USA). The range of effective angle of incidence (AOI) is from 38.8° to 58.8°. Both the AOI and optional polarization are automated with stepper motors, which provides for excellent electromechanical control. The control software changes the AOI and polarization and communicates with the FT-IR workstation software to perform both the background collection and spectrum collection with minimal effort. The entire system can produce multiangle and/or multipolarization ATR data sets of virtually any size. Each spectrum in the resulting data set has a corresponding background spectrum. The polarizer is removable. Polarized spectra have the polarizer in-line and unpolarized spectra have the polarizer removed. ATR-1-Plate-1: Zinc selenide (ZnSe) (20 mm diameter sampling area). ATR-1-Plate-2: Germanium (20 mm diameter sampling area). ATR-2: A PIKE GladiATR (Pike Technologies). ATR-2-Plate-1: Diamond (3 mm × 2 mm sampling area). ATR-2-Plate-2: Germanium (3 mm diameter sampling area). ATR-2 has a nominal AOI of 45° and does not have a polarizer.
Note that ATR-1 and ATR-2 have different focusing optics to support the much different areas of their ATR crystals. The AOI spread is comparably small and usually clearly below a standard deviation of 1° (Table S1, Supplemental Material). Accordingly, spectra can usually be successfully corrected as if there is no AOI spread at all. This even holds for standard deviations as high as 5° (Figures S22–S25, Supplemental Material). Note that such high standard deviations should be avoided for ZnSe and diamond ATR crystals and 45° AOI, as in this case for many materials a part of the light falls onto the sample under a subcritical angle so that the ATR condition is noticeably violated. This manifests itself in a reflectance in non-absorbing regions < 1 or, equivalently, in baselines > 0 in reflectance absorbance spectra.
Two similar FT-IR systems, designated FT-IR-1 and FT-IR-2, were used. FT-IR-1: A Thermo (Thermo Fisher Scientific) Nexus 670 hosting ATR-1. FT-IR-2: A Thermo Nexus 870 hosting ATR-2. The instruments sit side by side.
Chemicals
Consumer-grade distilled water was purchased from a retailer in Monona, Wisconsin, USA. The Octadecane (99%+) was purchased from Sigma-Aldrich.
Calibrations
There are two calibration methods used in this paper to obtain the polarization angle, a measurement of the net polarization of the ATR, and an FT-IR system in use. It is desirable to have a colorless sample with peaks in the MIR and a known
Calibration Method I
Calibration Method II
An ATR accessory with a polarizer does not require the water optical constants:
R(45°), Rs(45°), and Rp(45°) of water, or any suitable liquid, is recorded. The polarization angle is then determined with a fit based on Eq. 2.
Analysis
R(45°) is recorded on an ATR accessory without polarization.
Rs(45°) is calculated with Eq. 5 and the polarization angle from the calibration procedure.
The phase spectrum
With
Results and Discussion
To check our ATR correction formalism, we recorded s-, p-, and un-polarized ATR spectra of water on FT-IR-1-ATR-1-Plate-1. The experimental spectra are shown in Figure 1. We note that the relation derived from Fresnel's equations that

Experimental ATR spectra, recorded on FT-IR-1-ATR-1-Plate-1, of water with s- (black curve) and p-polarized incident light (green curve), as well as with unpolarized light (blue curve). For comparison, we also show the squared s-polarized spectrum (red spectrum, nearly completely covered by the green spectrum).
Usually the optical constant functions of water in combination with Fresnel's equations and Malus's law would be used to determine the polarization angle of the ATR accessory–instrument combination (Eq. 3) in combination with Eq. 4 (Calibration Method I), 15 we used the possibility provided by our accessory–instrument combination, which allows the employment of a polarizer, and determined the polarization angle directly for the sample by Eq. 4 (Calibration Method II). To that end, we recorded the s- and p-polarized spectra and computed the polarization angle by minimizing the residual sum of squared differences between the spectra recorded without a polarizer. By this procedure, we found that the polarization angle equals approximately 39.3°. As mentioned above, the (beam) spread is ±0.2°. Together these parameters are useful for establishing the suitability of the ATR accessory and optics for recording ATR correctable spectra of a given sample (see the table in the Supplemental Material). This will be discussed in more detail in a future paper. A comparison between the experimental and the fitted spectrum is provided in Figure 2.

Upper panel: Experimental ATR-spectra of water (recorded on FT-IR-1-ATR-1-Plate-1) with s- (black curve) and with unpolarized light (blue curve). The latter is compared with the simulated spectrum (red curve) according to Eq. 4 with the polarization angle of 37.5°. This angle was put into Eq. 5 to compute the calculated Rs (green curve). Lower panel: Differences between the experimental and simulated unpolarized spectrum as well as experimental and computed s-polarized spectrum.
Except for spectra below about 1000 cm–1, the experimental and the fitted spectrum are virtually congruent. A further check of consistency is the comparison of the s-polarized experimental spectrum and the one that can be computed from Eq. 5 by employing the determined polarization angle. Again, both spectra are virtually congruent. The remaining deviations are small enough to prove the value of Eq. 5, and, thereby, our fast and sophisticated ATR correction scheme. In this scheme, the last step is to apply the Kramers–Kronig analysis according to Eqs. 7 and 8.
Since there were only very small deviations between the squared s-polarized ATR spectrum and the p-polarized spectrum (Figure 1) it is clear that the combination of the sample, water, and the ATR-FT-IR optical system is sufficient to record R, Rs, and Rp spectra that are highly consistent with respect to the Fresnel equations and Malus’ law and this formalism can reliably use these data to calculate the complex refractive index function for the sample.
We can also expect that the differences between the complex refractive index functions determined from the s-polarized and the p-polarized ATR spectra are very small, since we have established that
In addition, since the differences between the experimental s-polarized ATR spectrum and the s-polarized ATR spectrum calculated from the spectrum gained without a polarizer based on Eq. 5 are also minor ones, the same can be expected for the deviations for the complex refractive index functions determined from these spectra. These expectations are fully confirmed by the results as illustrated in Figure 3, which shows the complex refractive index functions calculated from unpolarized spectra, as described in this paper, but also from s- and p-polarized spectra (how to obtain the complex refractive index function from p-polarized spectra shall be described in a future paper). This proves that our method of correction is not only fast, but also on the same level in this respect as other sophisticated methods such as iterative formalisms and dispersion analysis.2,17,21

While internally all results seem to be consistent, a comparison of the determined optical constants functions n(
Note that the correction scheme presented in this work can only work if the sample obeys Eq. 1 in the high-wavenumber limit, i.e., if the index of refraction n2(
Since dispersion is usually very small in the transparency region in between the MIR and the ultraviolet–visible (UV–Vis), the incorporation of n∞ assures that the errors of not extending the integral beyond the upper limit of the wavenumber range
One may object that the optical model assuming the ATR crystal as a semi-infinite incidence medium does not reflect the experimental conditions sufficiently well. In fact, Harrick
28
has shown that for s-polarized radiation the measured reflectance is given by:
Overall, one should keep in mind that there still are certain restrictions that might then require a more sophisticated method of correction such as dispersion analysis. 2 Finally, the formulas presented here (and elsewhere in the literature) do not allow us to evaluate p-polarized spectra. We will detail how we evaluated Rp, shown in Figure 3, in an extended future correspondence, where we will also cover measurements for different incidence angles.
Conclusion
Overall, we have developed a fast and sophisticated ATR-spectra correction scheme for spectra gained with unpolarized incident light and a 45° incidence angle. This scheme is based on a simple relation between s- and p-polarized reflectance which allows us to calculate the s-polarized spectrum from the spectrum recorded without using a polarizer if the polarization angle (in dependence of the wavenumber) is known. We have confirmed the validity of this scheme by experiment and by extensive discussion of potential limitations and errors. In our opinion, this sophisticated ATR correction scheme should replace the so-called advanced ATR correction scheme, which is in fact only more advanced than merely multiplying ATR spectra with the wavenumber, and which breaks down if the refractive index around a band causes the critical angle to become larger than the incidence angle.
Supplemental Material
sj-docx-1-asp-10.1177_00037028231219528 - Supplemental material for Sophisticated Attenuated Total Reflection Correction Within Seconds for Unpolarized Incident Light at 45°
Supplemental material, sj-docx-1-asp-10.1177_00037028231219528 for Sophisticated Attenuated Total Reflection Correction Within Seconds for Unpolarized Incident Light at 45° by Thomas G. Mayerhöfer, William D. P. Costa and Jürgen Popp in Applied Spectroscopy
Footnotes
Declaration of Conflicting Interests
The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding
The authors disclosed receipt of the following financial support for the research and/or authorship of this article: This work was supported by the EU, the Thüringer Ministerium für Wirtschaft, Wissenschaft und Digitale Gesellschaft, the Thüringer Aufbaubank, the Federal Ministry of Education and Research, Germany (BMBF), the German Science Foundation, the Fonds der Chemischen Industrie, and the Carl-Zeiss Foundation.
Supplemental Material
All supplemental material mentioned in the text is available in the online version of the journal.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.
