Abstract
Carbon dioxide production during cardiopulmonary bypass derives from both the aerobic metabolism and the buffering of lactic acid produced by tissues under anaerobic conditions. Therefore, carbon dioxide removal monitoring is an important measure of the adequacy of perfusion and oxygen delivery. However, routine monitoring of carbon dioxide removal is not widely applied. The present article reviews the main physiological and pathophysiological sources of carbon dioxide, the available techniques to assess carbon dioxide production and removal and the clinically relevant applications of carbon dioxide-related variables as markers of the adequacy of perfusion during cardiopulmonary bypass.
Introduction
Carbon dioxide (CO2) removal monitoring during cardiopulmonary bypass (CPB) is considered a recommended guideline for practice by the American Society of Extracorporeal Technology and a standard of practice by the Australian New Zealand College of Perfusion.1,2 This monitoring is usually performed through capnometric analysis of gases from the exhaust port of the oxygenator. Despite its relative easiness, this practice is not routinely applied; recent data are lacking in the literature, but a French survey in 2001 reported only 25% of the perfusionists using on-line capnography during CPB. 3
CO2 tension (PCO2) as measured by capnometry is representative of the complex interaction of CO2 production by cells and its elimination by the natural or artificial lung. Therefore, this technique may provide important information, not only on the efficacy of CO2 removal, but even on the metabolic status of the peripheral organs and on the adequacy of their perfusion with respect to their oxygen (O2) needs. 4 The present review article addresses the pathophysiological meaning of CO2 production during CPB, the technical aspects of CO2 kinesis measurement on CPB and the clinical relevance of this monitoring. Table 1 reports the nomenclature and abbreviations used throughout this article.
Nomenclature.
°: end tidal during normal or mechanical ventilation or at the exhaust port of the oxygenator during cardiopulmonary bypass.
Physiology and pathophysiology of CO2 production
The sources of CO2
Carbon dioxide and water are end products of energy production pathways. Carbohydrates and lipids are the substrates of the energy production; they are oxidized by O2 in aerobic conditions in the mitochondria, producing high-energy phosphate compounds and CO2. Six molecules of CO2 are produced and six molecules of O2 are consumed when the energy substrates are carbohydrates, with an RQ of 1.0; when lipids are oxidized, the ratio is 16 CO2 molecules produced per 23 O2 molecules consumed, with an RQ of 0.71.
An essential mechanism of this energy production pathway is the “proton shuttle” which carries protons from the cytosol into the mitochondria (Figure 1). Oxygen availability at the cellular level is the limiting factor of this process and, whenever the O2 flow to the mitochondria is inadequate, the NADH + H+ is re-oxidized to NAD through a different, so-called “anaerobic” pathway. This pathway is located at the level of the cytosol and is based on pyruvate re-oxidization of NADH + H+ and lactic acid production at a rate of two molecules per each glucose molecule undergoing anaerobic glycolysis.

Aerobic and anaerobic carbon dioxide (CO2) production. Oxidative mechanisms are responsible for aerobic CO2 production. Anaerobic energy production results in excess protons which are buffered to H2O and CO2, leading to anaerobic CO2 production.
Lactic acid is a strong acid with a pK of approximately 3.9, being totally dissociated at the normal pH of 7.4; buffering of lactic acid is essential in order to preserve the cell pH. The buffering mechanisms of lactic acidosis have been studied extensively in sport physiology and to a minor degree in critically ill patients. The commonly accepted model5–7 is shown in Figure 1. Bicarbonate is the main buffer of lactic acid, with intracellular production of CO2 that flows into the circulation to be finally eliminated by the lung. So, in the presence of anaerobic energy production, there is an “extra CO2” production on top of the aerobic CO2 production. For each mmol of lactic acid buffered, 1 mmol (22.3 mL) of extra CO2 is produced 8 and the total VCO2 results from VCO2 (aerobic) + VCO2 (extra, anaerobic). The bicarbonate buffering of lactic acid is not the only buffering mechanism. Actually, during exercise, it has been calculated that about 40% of lactic acid production is buffered by non-bicarbonate systems whereas bicarbonate-dependent buffers account for the remaining 60%. 9 Even if the theory of bicarbonate-based buffering of lactic acidosis has been challenged, 10 it remains accepted as the primary mechanism for extra CO2 production under hypoxic conditions in the critically ill patient.11,12
The two energy production pathways (aerobic – anaerobic) are not alternative, as they may simultaneously occur in energy production when the DO2 is inadequate to meet the oxygen needs (VO2). In turn, these conditions, which are labelled by hyperlactataemia and extra CO2 production may belong to different scenarios (Figure 2); on one side, the anaerobic energy production may be triggered by a high work rate during strenuous physical exercise when the energy demand exceeds the maximum aerobic energy delivery (DO2max). At the opposite side, in critically ill patients, the DO2 may be inadequate to support even the basic energy requirements (critical DO2) and, to preserve the metabolic needs of the peripheral organs, the anaerobic metabolism is activated and produces additional energy at the expense of lactate accumulation and extra CO2 production.

Anaerobic carbon dioxide (CO2) production in different scenarios. DO2: oxygen delivery; VCO2: CO2 production; VO2: oxygen consumption. Anaerobic CO2 production may be triggered by strenuous physical work, exceeding the energy production of the aerobic metabolism (right side of X-axis) or by a poor DO2, below the critical value needed to maintain the basic O2 needs (left side of X-axis).
Carbon dioxide production and its measurement during normal conditions and CPB
VCO2 assessment in natural lung-ventilating patients
CO2 production (VCO2) may be assessed according to the equation:
If VE (sweep gas during CPB) is expressed as L/min and exhaled CO2 as tension (mmHg), the equation becomes:
VCO2 (mL/min) = 1,000 x VE (L/min) x ePCO2 (mmHg)/760(mmHg), that is
This equation makes the assumption that the inspiratory CO2 fraction is negligible and that both VCO2 and VE are measured under standard temperature, pressure, dry (STPD) conditions. Conversion into body temperature, pressure, saturated (BTPS) conditions requires a conversion factor which varies according to the patient temperature and the ambient temperature and dryness. For a patient temperature of 32°C, an ambient temperature of 18°C and a water vapour pressure of 18 mmHg, the conversion coefficient is 0.874 and the final equation becomes:
The physiological concept underlying the above equations can be better understood considering that the FECO2 is directly related to the alveolar and arterial PCO2. Since the PaCO2 is a direct determinant of arterial pH, the equation may be seen as an expression of the goal of maintaining a stable pH. When the VCO2 increases, the ventilation must increase to maintain a stable pH, otherwise the pH will decrease (respiratory acidosis).
This approach to the relationship between VCO2, VE, and PaCO2 is actually correct in many physiological conditions. When the VO2 increases (as during physical exercise), the VE increases in a linear fashion to allow oxygen input and aerobically produced CO2 clearance. When the anaerobic energy production is triggered, then the VE increases at a higher rate to guarantee the additional clearance of excess CO2 produced through the anaerobic pathway (Figure 2). Measurement of FECO2 offers a reliable approximation of the PvCO2 and, therefore, of the peripheral organs’ (during exercise, basically muscles) VCO2 production, because the concomitant increase in the Qc guarantees an increased lung perfusion to match the increased lung ventilation.
However, things are different when the anaerobic CO2 production to buffer lactic acid is not physiologically coming from strenuous exercise, conversely being driven by an inadequate DO2 as it happens in the case of a low cardiac output (cardiogenic shock). In this clinical setting, the Qc is reduced and lung perfusion is consequently poor, with an overall increased pulmonary ventilation/perfusion ratio. This phenomenon is known as “venous blood CO2 stagnation” 13 and the excess CO2 produced to buffer lactic acid is accumulated on the venous side of the circulation, as reflected by an increase in the PvCO2. Conversely, given the low pulmonary perfusion, the CO2 elimination is reduced and the ePCO2 does not follow the PvCO2, creating the conditions for an ePCO2 – PvCO2 gradient. When this condition takes course, equation (1) provides a false underestimation of the VCO2 and cannot be applied.
Clinicians in charge of critically ill patients, where hyperlactatemia and the consequent excess CO2 production are common, are well aware of this problem and have searched for alternative measures of VCO2, considering this index as a valuable indicator of an inadequate DO2 due to a low Qc.
Unfortunately, assessing the VCO2 with equations not based on the measure of the FeCO2 is far from being an easy task. A theoretical alternative could be based on Fick’s principle, which is based on the assessment of the Qc and the arterial and mixed venous CO2 content:
Fick’s principle works nicely for the determination of the VO2 because the arterial and mixed venous O2 contents can be reliably calculated based on the pO2, the haemoglobin concentration and the haemoglobin O2 saturation, given the fact that the great majority of the O2 is carried by the haemoglobin.
Conversely, CO2 in blood is carried by different systems: dissolved CO2 (about 10%), bicarbonates (about 60%) and associated with proteins (about 30%). Therefore, a calculation of the CO2 content based only on the PCO2 (equation 2) is not possible, because the relationship between PCO2 and CO2 content is not linear (Figure 3) and varies depending on the pH status. Therefore, surrogates of VCO2 measurement have been proposed to assess cardiovascular failure. These include the measurement of the PvCO2 – PaCO2 gradient 11 based on the fact that, under conditions of stable VO2 and VCO2, this gradient increases along with the decrease of Qc.14,15 Alternative surrogates include the ratio between veno-arterial PCO2 difference and arterio-venous oxygen content. 16

Relationship between carbon dioxide (CO2) pressure and content. Acidosis decreases and alkalosis increases the CO2 content at the same PCO2.
VCO2 assessment in artificial lung ventilating patients (CPB)
The hollow-fibre/membrane oxygenators commonly used on CPB have totally different performance in terms of CO2 clearance with respect to the natural lung. Actually, the ventilation/perfusion ratio of the natural lung is around 0.8 (higher at the top and lower at the bottom of the lung in the orthostatic position). An increase of this ratio, as happens in the case of low pulmonary blood flow, creates the conditions for a PvCO2 – ePCO2 gradient. Conversely, artificial lungs have a much higher efficiency in clearing venous blood from the CO2, with an optimal ventilation/perfusion ratio of around 0.4-0.5. 17 As a consequence, the sweep gas mixture to the artificial lung rarely exceeds 50% of the Qc in the daily clinical practice in order to avoid an excessive CO2 removal.
This property of the artificial lung offers a great advantage for the measurement of VCO2 during CPB. The ability of the oxygenator to clear off the CO2, even at a low ventilation/perfusion ratio, avoids the onset of a PvCO2 – ePCO2 gradient and the ePCO2 measured at the exhaust port of the oxygenator can be reliably used to assess the VCO2 based on equation (1). Experimental studies during CPB have demonstrated that this calculation provides values of VCO2 that are associated with the correspondent VO2 at a higher degree than values obtained with Fick’s principle (equation 2). 18
Technical aspects of CO2 monitoring during CPB
The partial pressure of carbon dioxide (pCO2) may be measured by serial samples from the venous and arterial lines of the CPB circuit and modern monitoring devices allow a continuous, on-line measurement of both. However, due to the already mentioned non-linearity of the relationship between CO2 tension and content, these parameters cannot be used for a reliable VCO2 calculation.
Infra-red (IR) spectography is the most popular means currently used to monitor CO2. The wavelength of IR rays is >1.0 µm while that of visible light lies between 0.4 and 0.8 µm. Infra-red rays are absorbed by polyatomic gases (e.g. N2O and CO2), but water vapour also absorbs IR light. Carbon dioxide selectively absorbs specific wavelengths (4.3 µm) of IR light. Since the amount of light absorbed is proportional to the concentration of the absorbing molecules, the concentration can be determined by comparing the absorbance with that of a known standard. The CO2 concentration measured by the monitor is usually expressed as partial pressure in mmHg although some capnographs display percentage CO2 (FCO2) by dividing CO2 partial pressure by the atmospheric pressure. 19
Two types of capnograph are used for measuring CO2 in the gas exhaust of an oxygenator (ePCO2): mainstream and sidestream devices. In the mainstream devices, the CO2 sensor is mounted directly in the gas exhaust of the oxygenator. In order to prevent condensation of water vapour, which will give false high CO2 readings, mainstream sensors are heated above body temperature. The major advantages of mainstream devices are: no need for a sampling line, no risk for obstruction, not affected by changes in water vapour pressure, no delay in recording and can be used with low gas flows as encountered during neonatal and pediatric CPB. The main disadvantages are the weight of the sensor and the difficulty to insert the sensor into the oxygenator gas exhaust port which has typical a diameter of 1/4 inch or 3/8 inch. Sidestream capnographs have the CO2 sensor built into the console (away from the oxygenator) and a tiny pump aspirates gas samples from the oxygenator gas exhaust through long capillary tubing into the capnograph. The sampling tube is connected to a female Luer in the gas exhaust or to a large bore needle inserted in a piece of tubing connected to the gas exhaust port. The sampling flow rate can be set high or low. The latter is important because, if the ventilating gas flow is equal or lower than the sampling flow, room air can be aspirated and mixed with ventilating gas, resulting in artificially low ePCO2 values.
Most centres that use oxygenator exhaust capnography routinely use sidestream devices because of their ease of use. The exact position of the sample line connection in the gas exhaust is important as all microporous hollow-fibre membrane oxygenators are protected against accidental blockage of the gas exhaust port and the risk to transfer microemboli into the blood stream 20 by one or more vent ports. If the sampling is located too close to the vent ports, the sample can be diluted with room air. 21
The accuracy of IR spectroscopy during CPB can be influenced by the presence of water vapour. The difference in water vapour pressures between blood and gas will lead to condensation in the inner lumen of the gas fibres. To avoid interference in the measuring cell, as water vapour will absorb part of the IR rays, Nafion (a sulfonated tetrafluoroethylene-based fluoropolymer-copolymer) tubing in combination with a water trap should be used, virtually eliminating this source of analyzer failure.
Initially, capnography was mainly used to measure metabolic CO2 production and to estimate PaCO2. Many studies show a good correlation between ePCO2 and uncorrected PaCO2 during clinical cases,22-24 making it easier to maintain a specific target PaCO2 during CPB. Although there is a good linear correlation between ePCO2 and uncorrected PaCO2, there is some deviation, especially during the rewarming phase. This deviation is the same for oxygenators of the same type and brand, but can differ between types and brands. These differences are most likely due to differences in fibre bundle design, with countercurrent design having the best overall correspondence. 21
More recent research showed that VCO2 is a good predictor of anaerobic metabolism 25 and, as such, can help to reduce CPB-related morbidity. Indeed, CO2-derived parameters are more rapid and sensitive than O2-derived parameters in detecting anaerobic metabolism. 16 A combination of DO2, VO2 and VCO2 parameters was significantly associated with the risk of postoperative renal insufficiency. 26
Beside technical errors, clinical practice can also influence ePCO2 readings. In order to facilitate de-airing of cardiac cavities and to improve neurological outcome, surgical programmes start to use CO2 flooding of the operating field.27–29 However, the use of this technique has some major disadvantages as huge quantities of CO2 will enter the systemic circulation, leading to severe hypercarbia and acidosis during CPB. 30 As it is difficult to predict how much CO2 will enter the CPB circuit, it was suggested that capnography was used to monitoring ePCO2. 27 Under the guidance of ePCO2, the non-metabolic CO2 will be instantaneously detected and can be removed by increasing the oxygenator sweep gas flow, thus, maintaining a physiological acid-base status. However, it is important to understand that, when CO2 flooding is used, the ePCO2 values will no longer reflect metabolic CO2 production as it will be the combination of the metabolic CO2 production and the removed amount of non-metabolic CO2. This latter contribution may be limited, but not totally eliminated, using controlled CO2 flooding of the surgical field (i.e. not exceeding 1 L/min).
Clinical relevance of CO2 monitoring
The potential clinical relevance of VCO2 monitoring goes far beyond the safety control of the maintenance of an adequate PaCO2 of the patient throughout the CPB procedure. Certainly, the on-line measurement of ePCO2 at the exhaust port of the oxygenator may prompt sweep gas adjustments to rapidly adjust the systemic PaCO2 during the different phases of CPB. Increased values of ePCO2 as an expression of increased VCO2 can be found at the release of the cross-clamp, due to the reperfusion of the heart (anaerobic CO2) or during the rewarming phases after deep hypothermia, as an expression of the increasing VO2 (aerobic CO2) and of the decreased solubility of CO2. This last mechanism should be considered when assessing VCO2 during CPB cooling and rewarming phases, where the changes in CO2 solubility respectively decrease and increase the ePCO2. Nowadays, the routine use of low temperatures on CPB has been replaced by moderate hypothermia or normothermia in many institutions. However, CPB temperatures <28°C may still be used in congenital heart surgery and for specific interventions of high complexity. Within the setting of profound hypothermia, the changes in CO2 solubility result in corresponding changes in pH, with a reflection on cerebral blood flow. To compensate for the low values of CO2 at low temperatures, the pH-stat strategy considers the addition of exogenous CO2 to the sweep gas. Another exogenous source of CO2 may come from the flooding of the surgical field to prevent the formation of large air bubbles inside the heart chambers. In this case, adequate adjustments of the sweep gas may be prompted by the increase in ePCO2 due to the additional CO2 entering the venous line (exogenous CO2). The finding of elevated values of ePCO2 is rarely attributable to a failure of the oxygenator, which is much more often manifested by a poor blood oxygenation.
Apart from this, the measurement of the VCO2 may offer important information on the adequacy of the perfusion in terms of DO2 (pump flow x arterial O2 content). As outlined in the pathophysiology of CO2 production, under conditions of inadequate DO2, there is an excess CO2 production as a result of lactic acid buffering. Therefore, VCO2 may be considered as an indirect marker of lactate increase.
In 2006, Ranucci and associates 25 investigated the association between CO2-related parameters and arterial blood lactates in 54 adult patients during CPB. They found that the VCO2 and the DO2/VCO2 ratio were significantly associated with the arterial blood lactate concentration (Figure 4). With adequate statistical tests, cut-off values for the prediction of hyperlactataemia (>3 mmol/L) were found at a VCO2 of 60 mL/min/m2 and a DO2/VCO2 ratio of 5.0. The DO2/VCO2 was introduced to stress the role of an increased VCO2 due to a low DO2. Below the critical DO2, the VO2 becomes dependent on the DO2 and decreases. Simultaneously, aerobic CO2 production decreases and anaerobic CO2 production starts. However, anaerobic CO2 production only partially compensates for the decreased aerobic CO2 production 11 and, therefore, the absolute value of VCO2 may not be representative of a hypoxic peripheral tissue condition. Analogously, during the rewarming phases of CPB, the aerobic production of VCO2 is increased due to the increased VO2; however, the DO2 is increased as well by increasing the pump flow and the DO2/VCO2 ratio remains constant or, in any case, >5.0. Conversely, when the DO2 decreases below the critical threshold, the DO2/VCO2 ratio represents the combined effects of an inadequate DO2 and increased anaerobic CO2 production.

Panel A: Relationship between carbon dioxide production (VCO2) and arterial lactate concentration. Panel B: Relationship between oxygen delivery (DO2)/VCO2 ratio and arterial lactate concentration. Data from Ranucci and associates 24 during cardiopulmonary bypass (reproduced with permission).
In a recent study, 26 de Somer and associates confirmed the relevance of the DO2/VCO2 ratio on CPB as a predictor of postoperative acute kidney injury, with a cut-off value at 5.3, therefore, similar to the value of 5.0 identified as associated with hyperlactatemia. The use of DO2 and VCO2 to guide the perfusion management are now included in the concept of the “Goal Directed Perfusion (GDP)”.31–33 The GDP concept considers that the goal of perfusion is to maintain an adequate oxygen supply to all the organs, avoiding the patient entering into the anaerobic zone. To achieve this goal, the pump flow is adjusted according to the haematocrit value in order to reach and maintain a DO2 above the critical value of 270 mL/min/m2. 33 Within this context, the VCO2 should not exceed 60 mL/min/m2 and the DO2/VCO2 ratio should be higher than 5.0.
Of course, other measures to check the adequacy of perfusion with respect to the metabolic needs are available in daily clinical practice. The central venous oxygen saturation (SvO2) represents the O2 extraction from the peripheral organs and, therefore, increases when the DO2 decreases. However, during CPB, this measurement may be misleading: the SvO2 is measured at the level of the systemic venous return and represents a mixture of all the regional SvO2 values coming from the different organs. Since much of the venous return is coming from organs at a very low metabolic rate (i.e. muscles), the systemic SvO2 may appear normal, even if some peripheral organ (namely the splanchnic district) has a low regional SvO2. This concept has been demonstrated by McDaniel and associates 34 in an animal model, observing that the systemic SvO2 on CPB was normal despite very low values of regional SvO2 values from the cerebral and splanchnic districts. From a clinical perspective, a recent study demonstrated that acute kidney injury following cardiac surgery is associated with low levels of nadir DO2 on CPB, but not with the corresponding value of SvO2. 35 Additionally, another recent study found that the SvO2 values on CPB were not associated with the arterial blood lactate levels. 36
Arterial blood lactates are certainly a valuable measure of the adequacy of perfusion on CPB. Increased levels of arterial blood lactates on CPB have been associated with bad outcomes in both adult 37 and pediatric 38 heart surgery. However, the measurement of arterial blood lactates is not continuous and suffers from a number of limitations.
Once lactates are formed, they must be cleared by the liver and this process requires a variable, but considerable time. Therefore, a hyperlactataemia on CPB may reflect a condition of peripheral dysoxia occurring at any time before its detection and is not necessarily an expression of an inadequate perfusion. Additionally, after hyperlactatemia is established, blood lactates may remain high, even if the perfusion pattern is adequate.
A typical case is represented by patients under unstable haemodynamic conditions before the onset of CPB (Figure 5, a case from the series published by deSomer and associates 26 ). Hyperlactataemia is already present and the CO2-derived parameters are normal during the initial phases of CPB; a subsequent increase in arterial blood lactates due to inadequate perfusion is detected by the CO2-derived parameters.

A representative case of carbon dioxide production (VCO2) and oxygen delivery (DO2)/VCO2 ratio monitoring during cardiopulmonary bypass (CPB). Data from de Somer and associates. 26 Hyperlactataemia developed before the onset of CPB is not associated with increased VCO2 or decreased DO2/VCO2. New onset hyperlactataemia is detected by CO2-related parameters.
Conclusions
Despite the many possible applications of CO2-derived parameters during CPB, very few studies have been published in this area. The expanding concept of GDP will probably increase the interest of clinicians and researchers in these measurements. Further studies on the clinical relevance of CO2 production monitoring are warranted.
Footnotes
Declaration of Conflicting Interests
The authors declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: Marco Ranucci developed a patent for continuous DO2 and VCO2 monitoring on CPB. He is a consultant for Livanova Group and received honoraria from Medtronic Inc.
Funding
The authors received no financial support for the research, authorship and/or publication of this article.
