Abstract
Background
During cardiac surgery the use of a minimal extracorporeal circulation (MiECC) system may reduce the adverse effects for the patient. This is probably caused by reduced inflammation and hemodilution. For the use of a MiECC circuit, a venous bubble trap (VBT) is warranted for safety reasons. The aim of this study was to assess if an arterial filter with a small prime volume has the same (or better) air removal capacities as a VBT in a MiECC circuit and subsequentially may be used as an alternative.
Methods
In an in vitro study, air removal properties were compared between the arterial filter and three VBT’s on the market, VBT160 (Getinge), VBT 8 (LivaNova and VARD (Medtronic). In a MiECC circuit, the filter devices were placed in a venous position and challenged with massive and micro air. Gaseous microemboli (GME) were measured with a bubble counter proximal and distal of the VBT device.
Results
More than 99.9 % of the air was removed after a bolus air challenge by all VBT’s. Both the VARD and the AF100 showed better GME removal properties (not significant for the AF100) compared to the other devices. All filters showed GME generation after a challenge with massive air. Compared to the other filters, only the VARD showed no passing of larger bubbles when a volume of 50 mL of air was present in the filter.
Conclusions
The AF100 seems to be a safe and low prime alternative for use in a MiECC system as a venous air trap. A word of caution, placement of the AF100 arterial filter in the venous line is off label use.
Keywords
Introduction
During cardiac surgery the use of a minimal extracorporeal circulation (MiECC) circuit may reduce the adverse effects for the patient. A meta-analysis of Anastasiadis et al. 1 showed that the mortality rate of heart surgery was improved using MiECC, which is probably caused by reduced inflammatory response and less hemodilution.
Prime volume is the amount of volume needed to fill the extracorporeal circuit (ECC), typically 1.5-2 L, and is considered as a major contributor to hemodilution, because this volume is relatively large compared to the blood volume of the patient, which for an average patient is approximately 5 L. Ranucci et al. show in two studies that even a small reduction in prime volume may protect the patient from red blood cell transfusion 2 and may also affect goal directed perfusion strategies because hemoglobin levels are a major contributor to oxygen delivery to the patient. 3 In the latter study they show that an 1% decrease of nadir hematocrit during cardiopulmonary bypass (CPB) leads to a 7% increase in the relative risk of acute kidney injury. The reduced prime volume of a MiECC system (approximately 1 L) to a conventional ECC, especially in combination with autologous priming may contribute more to decreased hemodilution.
The reduction in hemodilution in a MiECC circuit is achieved by, amongst other techniques, the removal of the venous reservoir resulting in a closed circuit. For the use of a closed CPB circuit, a venous bubble trap (VBT) is warranted for air removal and safety reasons. 4 Air entering the MiECC system should be trapped by this filter and subsequently evacuated before reaching the arterial pump, the oxygenator, and, possibly, even the patient.
Characteristics of venous bubble filters included in this study and the AF100 arterial filter. The AF100 seems a safe alternative compared to devices which are specifically developed as a venous air filter.
(VARD = venous air removal device).
aThe VARD filter was used with the automatic air removal system.
The aim of this study was to assess if an arterial filter with a small prime volume has the same (or better) air removal capacities as a VBT in a MiECC circuit and subsequentially may be used as an alternative.
The air handling properties of the small prime volume arterial filter will be compared with the air handling properties of three other VBT’s.
Materials and methods
Design
A CPB test circuit was set-up based on a MiECC circuit and a pseudopatient (Figure 1). The MiECC circuit consisted sequentially of the tested air removal device in the venous line (Table 1), the VBT 160 (Bioline coated, Getinge, Rastatt, Germany), the VBT 8 (Phisio coated, LivaNova, Mirandola, Italy), the VARD (Venous Air Removal Device, Medtronic, Minneapolis, MN, USA) with the automatic air removal system (AAR1000 controller, Medtronic), or the Affinity AF100 (Medtronic). The circuit was then followed by a BPX80 Bio-pump (Medtronic) centrifugal pump and an Affinity Fusion oxygenator (Medtronic), with a gas blender setting of 2 L/min gas flow and with a fraction of delivered oxygen of 45%, resulting in a pO2 of 37-38 mmHg, followed by the pseudopatient. All Medtronic products were Balance® coated. The pseudopatient which functioned as a patient (volume) and as an air trap consisted sequentially of an AF100 (Medtronic) and an Affinity venous hardshell reservoir (Medtronic) which drained into the MiECC circuit. All tubing consisted of 3/8 inch and was Balance® (Medtronic) coated. Overview of MiECC circuit and pseudopatient study setup. The pseudopatient was used for all tests and consisted of an open venous hardshell reservoir and an AF100 filter, both for removing air from the circulation after the challenges. The location of the air filter is the position where the various venous bubble traps were placed. 1 and 2 are the positions of the bubble sensors. For further information see the ‘design’ section (Pvenous = pressuremeasurment in the venous line; MiECC = minimal extracorporeal circulation).
Bovine blood was obtained from the slaughterhouse and no animal research approval was needed. Bovine blood (diluted with Ionolyte (Fresenius Kabi, Huis ter Heide, Netherlands) to a hematocrit of approximately 25 %) was pumped with 4.5 L/min and flow was measured with an external flow probe (SonoTT, EMTEC GmbH, Finning, Germany). Blood temperature was maintained at 32°C with a heater (Julabo GmbH, Seelbach, Germany). Venous line pressure was adjusted with a Hoffman screw clamp to approximately −25 mmHg and the arterial line pressure was 125 mmHg.
Three sample devices of each brand of VBT and AF100 were tested.
Air challenges
The following air challenges were divided in bolus, gaseous microemboli (GME) and breakthrough challenges. The various in vitro challenges were setup to mimic, under controlled conditions, the different challenges which may occur in the clinical setting. The bolus and GME challenges are described earlier.6,7 A challenge was only started when the baseline showed a GME activity of <2 bubbles/s.
Bolus challenge
Using a 10 mL syringe on a three-way stopcock, 5 mL of air was quickly introduced (approximately 1 s) 40 cm proximal to the test VBT or AF100 (see Bolus challenge position in Figure 1). The collected air in the VBT or AF100 was removed within 5-10 s through the top purge port with a syringe. Air in the VARD filter was automatically removed with the Active Air Removal Controller (AAR1000, Medtronic). The controller was activated by the internal air sensors of the VARD.
The GME data were obtained from sensor 2 (see Figure 1 for position) and captured during a window of 60 s. This test was performed 5 times for each tested device.
Gaseous microemboli challenge
Air was introduced with a roller pump at 200 mL/min through ¼ inch tubing via a three-way stop cock in the venous line circa 40 cm proximal to the hardshell reservoir of the pseudopatient. This resulted in a continuous GME activity leaving the hardshell reservoir and challenging the tested filters. After reaching a constant GME load, GME was measured for 2 min. GME entering the device were measured with the bubble sensor 1 and the GME behind the device were measured with sensor 2 (see Figure 1 for the positions). From the measurement of both sensors the reduction rate could be calculated.
This test was conducted one time for each tested device. Meaning that the test was repeated three times, because three devices per brand were used for the study.
Breakthrough threshold
Using a 50 mL syringe placed on a three-way stop cock 40 cm proximal to the test device, a total of 50 mL of air was introduced in increments of 10 mL, 15 and 25 mL. The air trapped in the VBT or AF100 device was not evacuated. GME were measured continuously distal to the filter with sensor 2 (see Figure 1 for the position). This test was performed one time for each tested device. Meaning that the test was repeated three times, because three devices per brand were used for the study.
Gaseous micro emboli
GME were detected using the bubble counter BCC 300 (GAMPT mbH, Zappendorf, Germany). The measurements were conducted with two non-invasive sensor probes. The probes were clamped onto the venous line 5 cm distal to the venous hardshell reservoir and on the in-flow tubing of the oxygenator device (Figure 1). The BCC 300 accurately measures the number and the size of GME, with a diameter ranging from 20 to 2000 μm, using Doppler ultrasound (2 MHz). The device specifies the bubbles with a diameter greater than 2000 μm as ‘over range’.
Statistics
Quantitative variables are presented as mean ± standard deviation (SD) or as median with Interquartile ranges (IQR), when appropriate. We repeated five bolus tests for every tested device. Of these five tests the average was calculated and this value was considered as n = 1. The mean was calculated from the averages obtained from the same devices. Before analysis, data were tested for distribution according to Kolmogorov–Smirnov goodness of fit test. ANOVA or Kruskal–Wallis tests were used to assess statistical significant differences for the various groups. A p-value <.05 was considered significant. The Student’s t-test or Mann-Whitney tests were used, when appropriate, for comparisons focusing on two groups. Significance was reached according Bonferroni corrected p-values.
Statistical analysis was performed using SPSS 22.0 (SPSS, Chicago, IL, USA).
Results
Bolus challenge
The results of the bolus challenge of the various VBT and the AF100 devices are shown in Figure 2. An ANOVA test comparing the four groups showed a significant difference between the groups (p = .044). Comparing two groups showed a significant difference between the VARD filter (0.18 ± 0.07 µL) and both the VBT 8 and the VBT 160 (Volume: 0.88 ± 0.15 µL and 1.51 ± 0.44 µL; p = .002 and p = .007, respectively). No significant differences were found for pairwise comparison between the other devices, including the AF100 (1.11 ± 0.80 µL. Mean volume measured distal to the VBT or AF100 after bolus air challenge (n = 3). * significant difference compared to the VARD (VBT = venous bubble trap; VARD = venous air removal device).
Gaseous micro emboli challenge
The results of the GME challenge are shown in Figure 3. Between the four groups a significant difference in GME reduction was shown (ANOVA; p = .019). Testing between two groups showed significant differences between the VBT 8 and the VBT 160 (respectively, Volume: 14 ± 26% and 9 ± 25%; p = .008 and .006) compared to the VARD filter (88 ± 5%) only. No significant differences were detected between the AF100 (61 ± 40%) and the other devices. GME reduction (%) during GME challenge (n = 3). * significant difference compared to VARD (GME = gaseous microemboli; VARD = venous air removal device).
Breakthrough threshold
Observation of breakthrough threshold of every individually tested VBT or AF100 (n = 3) with cumulative air volume.
O = no GME detected distal to device; G = Continuous GME detection distal to device; B = Breakthrough (air larger than 2000 µm was observed).
(VBT = venous bubble trap; VARD = venous air removal device; GME = gaseous microemboli).

Example of an AF100 arterial filter after introduction of a total of 25 mL air during the breakthrough test. In this in vitro study setup with a bloodflow of 4.5 L/min, only GME were measured leaving the device (GME = gaseous microemboli).
Discussion
In this in vitro study an AF100 arterial filter was positioned in the venous line of a MiECC circuit and air removal properties were compared with three filter devices which were specifically developed as a venous air filter. The filters were challenged with various air challenges; macro air (bolus), micro air (GME) and massive air. We showed that the air removal properties of the AF100 arterial filter were good and comparable to the three VBT’s.
GME originate from the CPB/MiECC system and are caused by actions required during CPB, such as blood sampling, transfusing volume or injecting drugs.8,9 Besides, introduction of air into the venous line of the CPB circuit due to residual air in the venous cannula after connecting the tubes or by the excessive negative pressure will always lead to a detectable amount of GME in the arterial line.10–12
These GME are thought to contribute to the adverse outcomes of cardiac surgery. 13 The blood-air surface of GME can activate the inflammatory response, may damage the endothelium and obstruct the blood flow in the capillary vessels, causing transient ischaemia of end-organs. 14
A bolus air challenge was extremely well removed by all VBT’s and also by the arterial filter, AF100. A maximum of 1.5 µL was observed behind the tested filters after a bolus challenge of 5 mL, meaning that more than 99.97 % of the introduced air was removed. The VARD filter showed the best air removal properties of this bolus challenge. The automatic air removal probably added to this effect, since the air was quickly removed from the top of the filter therefore avoiding spilling of GME. In the other filters we choose to remove the collected air after approximately 10 s, which is equal to a clinical setting wherein the perfusionist manually removes air after observing visual air in the top of the filter. Large air volume in the filter will always lead to spilling of GME, as we showed in the breakthrough threshold test.
Interestingly, the filters without automated air removal showed similar results despite large differences in design. Air-removal by contemporary bubble filters is accomplished by a complicated combination of bubble buoyancy (influenced by velocity and centrifugal forces) and mechanical screen filtration. 15 It seems that reduction of larger air, as we did in the bolus challenge, was not dependent on pore size and volume of the filter. It is likely that the pressure necessary for the bubble to pass across the screen barrier, the bubble point pressure in our controlled study design, was not exceeded. 16
Both the VARD and the AF100 showed better GME removal properties (although not significant for the AF100) compared to the other devices. This is likely caused by the design of both filters. Both the VARD (without automated air removal device) and the AF100 were developed as an arterial filter with a small screen pore size (38 and 30 µm, respectively) and a larger surface area is created by folding of the filter material. This is also true for the VBT 8, however, this filter did not function better.
Detailed GME measurements in a previous study 4 showed that in MiECC system in a clinical setting a VBT 160, with a screen pore size of 160 µm, had a cut-off of approximately 400 µm, meaning that almost all GME larger than 400 µm were removed. We showed in another study that arterial filters with screen pore sizes between 25 and 40 µm have lower cut off values varying between 100 and 160 µm. 17 We may expect that the positioning of an arterial filter in the venous side of a MiECC system contributes more strongly to air removal which is supported by the data from our study. A word of caution, when massive air fills a filter with small screen pore size this filter is probably very difficult to deair.
We showed that the VBT 8 was the only filter spilling large bubbles at 25 mL of cumulative air. Both the VBT 160 and the AF100 started spilling larger air bubbles when 50 mL of air was present in the filter. Only the VARD showed no large over range bubbles distal to the filter. Possibly, as we also showed with the bolus challenge, the properties of the screen filter (i.e. surface tension, surface size and pore size), are the major contributors to avoiding air breakthrough.
Overall, the air removal properties of the AF100 arterial filter when placed in the venous line, are very good and comparable with devices which are specifically developed as a venous air filter. The AF100 seems to be a good GME scavenger, possibly reducing adverse effects of CPB. Additionally, the properties of removing or blocking massive air for the AF100 are similar to those of the venous filters, thus protecting the patient and adding to safety. According to ‘Guidance for Extracorporeal Blood Circuit Defoamer 510(k)’ of the FDA, in which testing for market approval of arterial and venous air traps are described, new devices should be equivalent in air removal compared to predicate devices. We believe that our study proves that the AF100 has equivalent air removal properties compared to predicate VBT’s. Accordingly, the AF100 seems to be a safe alternative, and this is achieved with the lowest prime volume. Since hemodilution is considered as a major contributor to the detrimental effects of CPB, the low prime volume of this filter can be considered as beneficial. Although the prime volume difference is low, especially in a CPB set-up for MiECC the AF100 may be a good option.
A limitation of this study may be that the air introduction was generated in vitro with a strict separation between a GME and a bolus challenge. The air challenge in a clinical setting may be a combination of both, which may influence the air removal properties of the devices.
We decided to measure behind the centrifugal pump, with the idea that this pump fractionates the bubbles and concordantly improving the measurement with the bubble counter. However, possibly the centrifugal pump also has air removal properties influencing the measured GME volume.
We compared the devices with volume measurements and did not specify distribution of the diameters. We know that the distribution was in the lower diameter range. GME volume measured after a bolus challenge was not higher than 1.5 µL, while a diameter of 2 mm (over range) will give a bubble of 4 µL. No over range bubbles were detected after the bolus and GME challenges.
Importantly, placement of the AF100 arterial filter in the venous line is off label use. According to the instructions for use by Medtronic the product should be positioned in the arterial line.
Conclusion
The air removal properties of the AF100 filter in the venous position are very good and comparable with devices which are specifically developed as a venous air filter. Both massive air and GME were scavenged equally well or better by the AF100 compared to the other filters.
In conclusion, the AF100 seems to be a safe and low prime alternative for use in a MiECC system as a venous air trap. A word of caution, placement of the AF100 arterial filter in the venous line is off label use.
Footnotes
Declaration of conflicting interests
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding
The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work was supported by an unrestricted grant of Medtronic.
