Abstract
Objective:
To determine whether the leaflets of bicuspid aortic valve (BAV) experience increased strain when compared to tricuspid aortic valve (TAV) leaflets.
Background:
The population at highest risk of aortic valve calcification (AVC) are individuals with BAVs. Currently, efforts to medically treat AVC are hampered by a limited understanding of the biomechanical forces involved in the molecular pathogenesis of AVC.
Methods:
Surgically created BAVs and control TAVs were placed into a left heart simulator. Strains were calculated by comparing the distances between points on the aortic valve (AoV) leaflet during various time points during a simulated cardiac cycle.
Results:
The fused leaflets of BAVs experience significantly more strain during systole when compared to TAVs. Specifically, BAVs experience 24% strain (P < .0001) in the radial direction, parallel to the direction of blood flow, as compared to TAVs. There was peak difference of 4% (P < .001) in the circumferential direction.
Discussion:
Based upon the data presented here, we are in the process of identifying how increased strain activates calcification-associated pathways in AoV cells. Future studies will examine whether these stretch responsive pathways can be blocked to inhibit calcification of BAVs.
Introduction
Aortic valve calcification/stenosis (AVC) is the third leading cause of adult heart disease 1 and the most common form of acquired valvular disease in developed countries. 2 The risk factor most closely linked to calcific aortic stenosis is bicuspid aortic valve (BAV), 2 –5 since 64% of the calcified aortic valves (AoVs) have BAV morphology. 6 Currently, the only treatment for severe AVC is replacement of the valve. Efforts to develop medical treatments for AVC have been hampered by a limited understanding of molecular mechanisms by which inflammatory and osteogenic pathways are activated in AVC.
The BAVs, the most common congenital heart defect with an incidence rate of 1% to 2%, 2,4,5 result from the failure of the aortic valve leaflets to separate during development. In BAVs, there is typically a fused aortic valve leaflet that is smaller than the two normal leaflets would have been. The other leaflet, the unfused leaflet, is typically of normal size. An ex vivo study of explanted BAVs from individuals who died from noncardiac causes suggests that the fused leaflet experiences significant stress. 7 This study did not quantify the biomechanical strains experienced by the fused leaflets nor compared them to the leaflets of a normal tricuspid aortic valve (TAV). Furthermore, attempts to quantity the changes in mechanical stretch experienced by BAVs and TAVs using computational models have been inconclusive. 8 Although much effort has been devoted to studying the role of the atherosclerotic processes in aortic valve (AoV) disease, there is evidence that aortic valve interstitial cells (AVICs) can calcify in a cell autonomous mechanisms. 9 –12 Although there have been recent reports examining the role of altered shear stresses in the leaflets of BAVs, 13 –15 very little is known about the strain experienced by BAVs. Elucidation of the hemodynamic forces imposed on the BAVs is important to the development of medical treatments for BAV calcific disease.
Methods
Quantification of Strain Experienced by BAVs and TAVs
Aortic valves were harvested from pigs sacrificed at the University of California San Diego Center for the Future of Surgery. The aortic valves and ascending aortic roots were mounted on Gore-Tex rings. To replicate the bicuspid AoV morphology, the left and right cusps were sutured together thereby replicating the most common valve morphology found in BAVs. For the purpose of quantifying the dynamic deformation and the resulting strain of the leaflets during the systolic phase of the cardiac cycle, we measured the relative displacement of small ink dots tattooed on the leaflets (each with an approximate diameter of 1 mm) during the opening phase of the valve. To maximize the resolution of the strain measurement, approximately 105 dots of an array of 15 × 7 were marked on a single leaflet. The valves were mounted in a replica of the Georgia Tech Left Heart Simulator (Figure 1). A controlled programmable pulsatile pump was used to simulate the ejection of the heart’s left ventricle. In order to simulate physiological conditions of a normal cardiovascular system, the peak cardiac output and heart rate were 35 mL/s and 70 beats/min, respectively. In addition, the downstream compliance chamber and the resistance were calibrated to retain the physiological conditions of a human heart. A mixture of 60% of deionized water and 40% of ethylene glycol was used to simulate the viscosity of blood of approximately 4 cP.

Experimental design. A, Pictures of aortic valves sutured to the Gore-Tex rings. The right picture is a bicuspid aortic valve (BAV). B, Diagram of the left heart stimulator used. C, Picture of the clear left ventricle stimulator used. D, Tracing of the flow pattern used to stimulate the blood flow through the aortic valve.
High-resolution cameras were placed at different angles to record the location of the dots marked on the leaflets throughout the opening of the valves. Strain was calculated by quantifying the change in distance between the centers of dots in either the radial or the circumferential direction. A time point during diastole was used as the reference point. We then transformed pairs of two-dimensional (2D) images acquired at two different angles onto three-dimensional surfaces at each time point. Once the instantaneous shape of the leaflet was reconstructed, we compared the initial and final distance of the dots in both radial and circumferential directions. Strain was calculated by comparing the change in distance between the points at the systolic time point being examined to the reference divided by the distance between points at the reference point. Three valves were used to test each condition.
Statistical Analysis
The statistical significance of differences between the groups was determined with the unpaired t test. P ≤ .05 was considered significant. Standard error of the mean was calculated and included in the graphs.
Results
The fused leaflet of BAVs experiences significantly more strain during systole. Given the lack of clarity regarding the relationship between valve morphology and the strain experienced by AoV leaflets, we performed ex vivo experiments using explanted porcine aortic valves. Based on the observation (Figure 2A and Supplemental video), the valves of both morphology closed and deformed in a similar manner at diastole. However, during systole, the BAVs deformed very differently than the TAVs most notably during peak systole. As a result of these observations, we focused on quantifying the strains experienced by the valves during systole. End diastole was used as the reference point, since TAVs and BAVs were similar at this time point. Although there was a statistically significant difference (P = .001), the magnitude of the peak strains in circumferential direction was very similar between the BAVs (14%) and the TAVs (10%). During mid-to-late systole, the fused valve leaflets of BAVs experienced a significantly higher peak strain of 24% in the radial stretching when fully open as compared to the peak of 4% in the TAV (P < .0001 at the measured time points; Figure 2D and Table 1).

Bicuspid aortic valves (BAVs) experience significantly more strain during systole. A, Still frames of the aortic valve leaflets during the cardiac cycle. The left frames are at diastole (the reference point). The right images are at peak systole. B, An image of a tricuspid aortic valve (TAV) with the ink dots on the ventricular side of the aortic valve (AoV) leaflet. The blue arrow demarcates the radial direction (parallel to blood flow). Radial direction is parallel to blood flow while the circumferential direction parallel to the aortic valve annulus plane. C, There is a 4% increase in the strain of BAV in the circumferential as compared to the TAV. D, The BAVs experience peak difference of 24% more strain in the radial direction as compared to the TAVs. N = 3 different valves for each group.
Quantification of Stretch. Average stretch values were computed at 6 different time points from early systole to late systole in both radial and circumferential direction. Stretch values were computed as the changes of distance of dots divided by the distance of reference. N = 3 different valves for each group.
Discussion
In this report, we demonstrate that BAV leaflets experience significantly more radial strain during systole as compared to TAVs. This is the first report experimentally comparing the strains experienced by BAVs as compared to TAVs during systole. We have focused on the strain imposed on BAV leaflets, since strain is of much greater magnitude than shear stress. Additionally, the strain directly affects AVICs, which are the cells that calcify in AVC. Previously, Robicsek et al examined the strain experienced by explanted human BAVs; however, they did not compare the differences between BAVs and TAVs. 7 On the other hand, computer modeling predicted minimal changes in the forces experienced by leaflets in TAVs and BAVs. 8 Our data demonstrate that the fused leaflets of BAVs experience significantly more strain in the radial direction than TAV leaflets. One possible explanation for the difference between the experimental data and the in silico 8 data is that the in silico BAVs were modeled as having symmetric valve leaflets instead of the morphology seen in patients in which the fused leaflet is larger than the unfused one. A limitation of our study is that we used surgically created porcine BAVs. This approach was necessary since large animals with BAVs and noncalcified human BAVs were not available. Based upon our data, we believe that the fused leaflets of BAVs experience significantly more strain than the leaflets in normal TAVs. Additionally, we propose that the patients with BAVs that experience increased strain will be the subset of developing AVC. Future efforts to noninvasively measure the valve leaflet strains would be a powerful tool for risk stratification patients with BAV for their potential of developing significant AVC.
It has been reported that cultured aortic valve, AVICs exposed to cyclic stretch have an increased expression of genes associated with AVC. 16,17 In separate reports, we are in the process of describing the discovery that stretch-mediated repression of a microRNA and a long noncoding RNA are critical to the increased expression of calcification-associated genes in human AVICs. We have previously demonstrated that microRNAs are altered in BAVs. 18 Based upon the data presented here and the aforementioned molecular data from our laboratory, we have developed the model that stretch activation of calcification-associated genes in response to the increased strain imposed on AVICs in BAVs is sufficient to cause AVC (Figure 3). This model will serve as a roadmap to identifying stretch responsive pathways that can be targeted to prevent calcification of BAVs. Ongoing efforts to elucidate the role that biomechanical stimuli plays in the pathogenesis of calcific disease of BAVs are important development of noninvasive treatments for this serious disease.

Model of how biomechanical stretch activate calcification related pathways in bicuspid aortic valves (BAVs).
Footnotes
Acknowledgment
The authors would like to thank the Center for the Future of Surgery at University of California San Diego for their assistance in obtaining the porcine aortic valves.
Declaration of Conflicting Interests
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding
The author(s) disclosed receipt of the following financial support for the research, authorship and/or publication of this article: VN was supported by NIH K08 HL086775-01.
Supplemental Material
Abbreviations and Acronyms
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.